VIII: MaintenanceDisposalCAUTION• Only service technicians properly trained by NIDEK can repair the device.NIDEK is not responsible for any accidents

Seite 3

86OPERATING PROCEDURES: Parameter Settings[FUNCTION 1 (Various functions 1)]* Underlined options indicate factory settings61 : WINDOW CHECK (measuring

Seite 4 - Use precautions

87OPERATING PROCEDURES: Parameter Settings266 : AR BRIGHTNESS (light intensity of AR measurement screen)Brightness of the LCD screen during the ARK me

Seite 5

88OPERATING PROCEDURES: Parameter Settings73 : CHANGE SPEED (Measurement changeover speed)Setting of the time period required when the measurement mod

Seite 6 - During Use

89OPERATING PROCEDURES: Parameter Settings22.8.2 Setting the date and timeWhen the date and time of the printout is not correct, set the correct date

Seite 7

90OPERATING PROCEDURES: Parameter Settings4 Press the up button or down button tomove the highlight to an item to be changed.5 Press the right but

Seite 8 - Wavelength (nm)

91OPERATING PROCEDURES: Parameter Settings22.8.3 Entering commentsComments to be printed can be changed (factory setting: “NIDEK TONOREF II”).1 Press

Seite 9 - After Use

92OPERATING PROCEDURES: Parameter Settings4 Press the right button or left button toselect the position where a character is enteredor changedThe

Seite 10 - Disposal

9333.OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTEDThe TONOREF II exports data to an external device such as NIDEK motorized refractor (hereafterrefe

Seite 11 - { Patient environment

94OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED: Connecting to the NIDEK Motorized Refractor (RT) or Computer3.1.2 Connecting procedure1 Connect th

Seite 12 - Table of Contents

95OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED: Connecting to the NIDEK Auto Lensmeter33.2 Connecting to the NIDEK Auto Lensmeter (LM)3.2.1 Outline

Seite 13

IX:{ Patient environmentThe patient environment represents a space where there is a possibility of direct contact between thepatient or the operator a

Seite 14

96OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED: Connecting to the NIDEK Auto Lensmeter (LM)3.2.3 Operating procedure1 After lens measurement with

Seite 15 - BEFORE USE

97OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED: Connecting to the Eye Care Card System33.3 Connecting to the Eye Care Card System3.3.1 OutlineData

Seite 16 - 1.2 Indications for Use

98OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED: Connecting to the Eye Care Card System3.3.3 Transferring data with the EyeCa-RWThe procedure for wr

Seite 17 - 1.3 Principles

99OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED: Connecting to the Eye Care Card System3{ When not printing measured dataSet the 39. PRINT parameter

Seite 18 - 1.4 Device Description

100OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED: Connecting to the Eye Care Card System

Seite 19

10144.MAINTENANCE4.1 TroubleshootingIn the event that the device does not work correctly, correct the problem according to the followingtable before c

Seite 20 - { Rear view

102MAINTENANCE: TroubleshootingIf the symptom cannot be corrected with the above actions, contact NIDEK or your authorized distributor.The auto-tracki

Seite 21

103MAINTENANCE: Error Messages and Countermeasures44.2 Error Messages and CountermeasuresIf one of the following error codes is displayed on the scree

Seite 22 - { Bottom view

104MAINTENANCE: Error Messages and CountermeasuresERR021 • Error related to control signals for communication (input port)Ensure that the interface ca

Seite 23 - 1.5.1 R/K measurement screen

105MAINTENANCE: Error Messages and Countermeasures4ERR051 • Error related to the movable prism.A malfunction of the movable prism motor, movable prism

Seite 24

Table of ContentsX1. BEFORE USE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 Outline of the device. . . . . . . . . . .

Seite 25 - ǁᵃᵆᵋᵆᵐᵅǂ

106MAINTENANCE: Replacing Printer Paper4.3 Replacing Printer PaperWhen a red line appears on the side of printer paper, it means that paper is running

Seite 26

107MAINTENANCE: Replacing Printer Paper43 Insert new printer paper.Load printer paper as shown in the picture on theright.Set printer paper so that it

Seite 27 - Parameter button

108MAINTENANCE: Fixing Chinrest Paper4.4 Fixing Chinrest Paper1 Disconnect the two fixing pins from the chinrest.2 Remove a proper number of chinrest

Seite 28 - 1.5.2 NT measurement screen

109MAINTENANCE: Checking the AR/KM Measurement Accuracy44.5 Checking the AR/KM Measurement AccuracyTo check the accuracy of measured data, use the pro

Seite 29 - ÁÖ

110MAINTENANCE: Checking the AR/KM Measurement Accuracy{ Values marked on the labels of the spherical model eyeVertex distanceDiopter (Unit: D)Corneal

Seite 30

111MAINTENANCE: Cleaning44.6 CleaningWhen the cover or panel of the device becomes dirty, wipe with a soft cloth. For stubborn dirt, immersethe cloth

Seite 31 - [Underside view]

112MAINTENANCE: Cleaning5 Check if the window is cleaned using a penlight. If not, clean it again with new cleaningpaper.Apply light with a penlight a

Seite 32 - 1.7 Checking Contents

113MAINTENANCE: Cleaning44.6.2 Cleaning the air nozzle1 Display the NT measuring screen.2 Check the glass part of the air nozzle from an oblique dire

Seite 33 - 1.8 Before First Use

114MAINTENANCE: List of Replacement Parts4.6.3 Cleaning the printerAfter repeated usage, the paper slot of the auto cutter of the printer may become s

Seite 34 - 9 Set the printer paper

11555.SPECIFICATIONS AND ACCESSORIES5.1 Classifications[Classification under the provision of 93/42EEC (MDD)] Class IIaThe TONOREF II is classified as

Seite 35 - { Setting by parameter

XI:2.8 Parameter Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 762.8.1 Parameter tab

Seite 36 - BEFORE USE: Before First Use

116SPECIFICATIONS AND ACCESSORIES: Safety Features5.2 Safety FeaturesTo ensure safe use, the device is provided with the following safety features.<

Seite 37 - OPERATING PROCEDURES

117SPECIFICATIONS AND ACCESSORIES: Specifications55.3 Specifications{ Measurement of refractive error (AR measurement){ Measurement of corneal curvatu

Seite 38 - Turn the power switch on ( )

118SPECIFICATIONS AND ACCESSORIES: Specifications{ PD measurement{ CS measurement{ PS measurement{ Measurement of intraocular pressure (NT measurement

Seite 39 - 3 Perform checks before use

119SPECIFICATIONS AND ACCESSORIES: Specifications5{ Other functions{ Dimensions and weight{ Environmental conditions (during use){ Environmental condi

Seite 40 - 5 Prepare the patient

120SPECIFICATIONS AND ACCESSORIES: Standard Configuration5.4 Standard Configuration5.4.1 Standard accessories5.4.2 Optional accessories• Printer paper

Seite 41

12166.EMC (ELECTROMAGNETIC COMPATIBILITY)The Electromagnetic Compatibility Directive sets the essential requirements for electrical andelectronic equi

Seite 42 - 7 Print the measured results

122EMC (ELECTROMAGNETIC COMPATIBILITY)Guidance and manufacturer's declaration - electromagnetic immunityThe TONOREF II is intended for use in the

Seite 43

123EMC (ELECTROMAGNETIC COMPATIBILITY)6Guidance and manufacturer's declaration - electromagnetic immunityThe TONOREF II is intended for use in th

Seite 44 - 㨃㧵㧺㧰㧻㨃ޓ㧯㧴㧱㧯㧷ޓ㧻㧷㧍

124EMC (ELECTROMAGNETIC COMPATIBILITY)Recommended separation distances between portable and mobile RF communications equipment and the TONOREF IIThe T

Seite 45

12577.GLOSSARY{ Terminology related to the AR/KM measurementz AI modeIn this mode, the measurement is automatically completed after three or more meas

Seite 46

: XII5.4 Standard Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1205.4.1 Standard access

Seite 47 - 2.3.1 Normal shutoff

126GLOSSARY: z Focusing indicatorThe indicator which shows the distance between the corneal center of the patient’s eye andthe tip of the air nozzle.z

Seite 48

127GLOSSARY:7z Vertex Distance (V.D.)The distance between the corneal vertex to the posterior surface of the spectacle lens.{ Terminology related to t

Seite 49 - 2.4 Selecting the Mode

128GLOSSARY: z Low confidence dataMeasured data with a “*” indication. This is displayed on the screen when measurement isperformed in spite of a meas

Seite 50

12988.INDEXAAir nozzle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Alignment spot . . . . . . . . . . . . . . . . . . .

Seite 51

130INDEX:

Seite 52 - 2.4.2 Sleep mode

111.BEFORE USE1.1 Outline of the deviceAUTO REF/KERATO/TONOMETER Model TONOREF II is designed to singly perform objectiverefraction, corneal shape mea

Seite 53 - Measurements

2BEFORE USE: Indications for Use1.2 Indications for UseThe AUTO REF/KERATO/TONOMETER TONOREF II is a medical apparatus which performs mea-surement of

Seite 54

3BEFORE USE: Principles11.3 Principles1. Objective refractionFine measurement beams are projected on the fundus of the patient’s eye by a projectingop

Seite 55

4BEFORE USE: Device Description1.4 Device Description{ Front viewFunction buttonsUsed to set the device and to switch the screen.Functions assigned to

Seite 56

5BEFORE USE: Device Description1• CLR button ( )Used to clear the measured data.When the CLR button is pressed for about a second, all the measured d

Seite 57 - 4 Measurement starts

NIDEK CO., LTD. : 34-14, Maehama, Hiroishi-cho, Gamagori, Aichi 443-0038, Japan(Manufacturer) Telephone: (81-533) 67-6611Facsimile: (81-533) 67-6610NI

Seite 58

6BEFORE USE: Device Description{ Rear viewForehead restDuring measurements, the patient’s forehead should be gently placed over the forehead rest.Clea

Seite 59 - 5 Measurement completes

7BEFORE USE: Device Description1Chinrest up/down buttons ( , )Move up or down the chinrest.Safety stopperUsed to provide a safety space so that the a

Seite 60 - 9 Print the measured results

8BEFORE USE: Device Description{ Bottom viewPower inletUsed to connect the detachable power cord.RS-232C connectorConnect an interface cable to send/r

Seite 61 - { CATARACT measurement mode

9BEFORE USE: Measurement Screen Description11.5 Measurement Screen Description1.5.1 R/K measurement screenThe screen for the AR (Refractive error) and

Seite 62

10BEFORE USE: Measurement Screen DescriptionAuto button ( )Selects the auto-tracking function and auto-shooting function.Select auto-tracking from 3D,

Seite 63

11BEFORE USE: Measurement Screen Description1Focusing indicatorIndicates the distance between the main body and the patient’s eye.Operate the joystick

Seite 64 - 2 Measurement completes

12BEFORE USE: Measurement Screen Description<When Page 2 is displayed>* CS, PS, and PD data, and cataract measurement mode mark are displayed on

Seite 65 - KM Measurement Mode

13BEFORE USE: Measurement Screen Description1Eyeprint button ( )Prints the eyeprint view of measured data.The eyeprint is printed out regardless of it

Seite 66

14BEFORE USE: Measurement Screen Description1.5.2 NT measurement screenThe following is the screen for NT (Tonometry) measurement.<When Page 1 is d

Seite 67 - ᵐᵑᴾᵏᵑǽᵐᵔǽᴷǽᵀᴾᵍᵑᵒᵏᵂ

15BEFORE USE: Measurement Screen Description1Eyelid detection mode button ( )Used to activate the detection (eyelid detection) mode which detects whet

Seite 68

I:Use this device properly and safely."CAUTION! Federal Law (US) restricts this device to sale by or on the order of a physician ora properly li

Seite 69 - PS (Pupil Size) Measurement

16BEFORE USE: Labels and Indications on the Device1.6 Labels and Indications on the DeviceTo call the operator’s attention, the device is provided wit

Seite 70 - ᴹᵍᵒᵍᵆᵉǽᵐᵆᵗᵂᴻ

17BEFORE USE: Labels and Indications on the Device1[Underside view]

Seite 71 - Manual PD Measurement

18BEFORE USE: Checking Contents1.7 Checking ContentsUnpack the contents from the shipping carton and check them.The following are included in the stan

Seite 72 - to exit from PD mea

19BEFORE USE: Before First Use11.8 Before First UsePlace the device on a stable table and connect a power cord to it.1 Place the main body on a stabl

Seite 73 - Contact lens

20BEFORE USE: Before First Use7 Turn the power switch on ( ).The initial screen is displayed on the LCD dis-play and the device starts initializing.8

Seite 74

21BEFORE USE: Before First Use1{ Please see here when you want to do like this.{ Setting by parameterSetting parameters allows various functions of th

Seite 75

22BEFORE USE: Before First UseSetting contents ParametersMeasurement count of NT measurement 25. NT CONTINUEWhether NT measurement values are displaye

Seite 76

2322.OPERATING PROCEDURES2.1 Operation Flow2.2 Preparation for Measurement (page 24)Turn on the device and configure it us as necessary.Set up the pat

Seite 77 - ÁÖ

24OPERATING PROCEDURES: Preparation for Measurement2.2 Preparation for Measurement1 Turn the power switch on ( ).The title screen is displayed and the

Seite 78

25OPERATING PROCEDURES: Preparation for Measurement22 The measurement screen is displayed.The measurement screen with the measure-ment mode (R/K or NT

Seite 79

II: Use precautionsBefore UseCAUTION• Do not use the device for other than the intended purpose.NIDEK is not responsible for accidents or malfunctions

Seite 80 - 8 Start the measurement

26OPERATING PROCEDURES: Preparation for Measurement3: CYL modeCylinder mode, the reading direction of cylinder data in which CYL data (cylindrical pow

Seite 81

27OPERATING PROCEDURES: Preparation for Measurement24) Adjust the height of the chinrest by the chin-rest up/down button ( , ) until the cen-ter level

Seite 82

28OPERATING PROCEDURES: Preparation for Measurement7 Print the measured results.Printing operation varies according to the 31. PRINT parameter setting

Seite 83

29OPERATING PROCEDURES: Preparation for Measurement22.2.1 Measuring window check for soiling and puffed air pressure check during startupIt is possibl

Seite 84 - 2.6.1 Eyelid detection mode

30OPERATING PROCEDURES: Preparation for Measurement2) The check result is displayed.• “PRESSURE TEST OK” is displayed:The air nozzle is clean.• If on

Seite 85 - 2.7 Printing

31OPERATING PROCEDURES: Preparation for Measurement26) At the completion of the checks, the screen returns to the measurement screen.2.2.2 Switching b

Seite 86 - <Sample printout 2>

32OPERATING PROCEDURES: Preparation for MeasurementDuring the switching process, the“ ” (during the switchingprocess from R/K mode to NT mode) or“ ” m

Seite 87

33OPERATING PROCEDURES: Finishing the Measurements22.3 Finishing the Measurements2.3.1 Normal shutoff1 Turn off ( ) the power switch.It is allowed to

Seite 88 - 2.7.2 Eyeprint

34OPERATING PROCEDURES: Finishing the Measurements4 When the “PACKING POSITION IS COM-PLETED / SHUT DOWN PLEASE” message isdisplayed, turn the power s

Seite 89 - 2 Press the print button

35OPERATING PROCEDURES: Selecting the Mode22.4 Selecting the ModeSelect the measurement mode from the following options:{ Measurement itemsSelect the

Seite 90 - 2.8 Parameter Settings

III:CAUTION• Do not use a power cord other than the one supplied. Also do not connect thesupplied power cord to any other device.Failure or fire may r

Seite 91

36OPERATING PROCEDURES: Selecting the ModeAR/KM measurement mode is selected by default; there is no need to select when proceed-ing to measurements i

Seite 92 - { Resetting the parameters

37OPERATING PROCEDURES: Selecting the Mode22.4.1 Switching to manual modeBy pressing the manual mode button during measurements, it is possible to t

Seite 93 - 2.8.1 Parameter tables

38OPERATING PROCEDURES: Selecting the Mode2.4.2 Sleep modeThe device goes into sleep mode automatically to save power if no button have been pressed f

Seite 94 - [NT (NT measurement)]

39OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements22.5 AR (refractive error) and KM (corneal curvature radiu

Seite 95

40OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements1 Instruct the patient to look through the measuringwindo

Seite 96 - [PRINT1 (Print setting 1)]

41OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements23D auto-tracking1) Perform rough alignment and focusing b

Seite 97 - [PRINT2 (Print setting 2)]

42OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) MeasurementsWhen the main body is not within the working range of auto

Seite 98

43OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements24 Measurement starts.Measurements take place automaticall

Seite 99 - [PRINT3 (Print setting 3)]

44OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements<AR measurement error messages><KM measurement er

Seite 100

45OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements25 Measurement completes.• When AI mode is YES:*1When the

Seite 101

IV: During UseWARNING• Before starting NT measurement, set the safety stopper for each patient to prevent theair nozzle from touching the patient’s ey

Seite 102

46OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements6 Measure the other eye in the same manner.When the other

Seite 103 - 㧞㧠ޓ㧝㧣㧦㧜㧜

47OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements2{ CATARACT measurement modeIf cataract or abnormal eyes c

Seite 104

48OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements{ Measurement ring image displayThe TONOREF II projects me

Seite 105 - 2.8.3 Entering comments

49OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements22.5.2AR (refractive error) Measurement: AR Measurement Mo

Seite 106

50OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements<AR measurement error messages>Error messages are th

Seite 107 - 3.1.1 Outline

51OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements22.5.3 KM (corneal curvature radius) Measurement: KM Measu

Seite 108 - 3.1.3 Operating procedure

52OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements<KM measurement error messages>Error messages are th

Seite 109 - 3.2.2 Connecting procedure

53OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements22.5.4 CS (Corneal Size) Measurement1 Press the CS/PS/PD

Seite 110 - 3.2.3 Operating procedure

54OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements6 Press the right button or left button toalign the gu

Seite 111 - 3.3.2 Method of connection

55OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements22.5.5 PS (Pupil Size) MeasurementThis is the procedure to

Seite 112 - 3 Press the print button

V:CAUTION• There may be a few constantly-lit, missing, or dead pixels in your LCD which are acharacteristic of the LCD. This does not represent failur

Seite 113 - 2 Insert the Eye Care card

56OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements5 Press the right button or left button toalign the gu

Seite 114

57OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements22.5.6 PD (Pupillary Distance) Measurement{ Auto-PD measur

Seite 115 - MAINTENANCE

58OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements3 After proper alignment of the right eye and lefteye, pre

Seite 116

59OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements22.5.7 Measuring Hard Contact LensesTo measure hard contac

Seite 117

60OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements4 Place the model eye/CL holder with the sur-face of the c

Seite 118

61OPERATING PROCEDURES: NT (Tonometry) Measurement: NT Mode22.6 NT (Tonometry) Measurement: NT ModeDuring NT measurement, the types of the target and

Seite 119

62OPERATING PROCEDURES: NT (Tonometry) Measurement: NT Mode3 Set the safety space between the patient’s eye and air nozzle with the safety stop-per.1)

Seite 120 - 4.3 Replacing Printer Paper

63OPERATING PROCEDURES: NT (Tonometry) Measurement: NT Mode25 Manipulate the joystick to display the patient’s eye on the screen.By tilting the joysti

Seite 121 - 3 Insert new printer paper

64OPERATING PROCEDURES: NT (Tonometry) Measurement: NT ModeIn the case of 3D auto-tracking:1) Perform rough alignment and focusing bymanipulating the

Seite 122 - 4.4 Fixing Chinrest Paper

65OPERATING PROCEDURES: NT (Tonometry) Measurement: NT Mode2If the main body gets out of the working range of auto-racking in the back-and-forth direc

Seite 123

VI: CAUTION• Information on the avoidance of overexposure to potentially hazardous opticalradiation (ISO 15004: 1997)Spectrally weighted photochemical

Seite 124

66OPERATING PROCEDURES: NT (Tonometry) Measurement: NT Mode8 Start the measurement.• When the auto-shooting function is on:The measurement takes place

Seite 125 - 4.6 Cleaning

67OPERATING PROCEDURES: NT (Tonometry) Measurement: NT Mode2<Error messages during NT measurement>• If the measurement was not performed correct

Seite 126

68OPERATING PROCEDURES: NT (Tonometry) Measurement: NT Mode{ If the “CHECK THE EYE” error occurs:Check the condition of the patient’s eye.If the patie

Seite 127 - 4.6.2 Cleaning the air nozzle

69OPERATING PROCEDURES: NT (Tonometry) Measurement: NT Mode211Pull the joystick backward once and move themain body to the other eye.The measured eye

Seite 128 - 4.7 List of Replacement Parts

70OPERATING PROCEDURES: NT (Tonometry) Measurement: NT Mode2.6.1 Eyelid detection modeWith this mode, the device always checks the eye for the amount

Seite 129 - 5.1 Classifications

71OPERATING PROCEDURES: Printing22.7 Printing2.7.1 Printing measured dataMeasured data is printed out by pressing the print button after measurement

Seite 130 - 5.2 Safety Features

72OPERATING PROCEDURES: Printing<Sample printout 2>• Items marked with “ ” indicate whether or not to print according to parameter settings.See

Seite 131 - 5.3 Specifications

73OPERATING PROCEDURES: Printing2* 1 Vertex distanceThe distance between the corneal vertex to the posterior surface of spectacle lenses.* 2 Near work

Seite 132

74OPERATING PROCEDURES: Printing* 10 PD for near visionPD for a near working distance of 40 cm (factory setting). Use it for prescriptions of reading

Seite 133 - { Others

75OPERATING PROCEDURES: Printing22.7.3 Printing parameter settingsThe present parameter settings, set time, comments and maintenance program versions

Seite 134 - 5.4 Standard Configuration

VII:After UseCAUTION• When the device is not in use, turn off the power switch and put the dust cover overthe device.If not, dust may affect the measu

Seite 135 - EMC (IEC 60601-1-2: 2001)

76OPERATING PROCEDURES: Parameter Settings2.8 Parameter SettingsThe TONOREF II is provided with parameters that set various functions according to the

Seite 136

77OPERATING PROCEDURES: Parameter Settings23 Press the execute button to switch to the[PRINT1] screen.Parameters and their settings are displayed.A

Seite 137

78OPERATING PROCEDURES: Parameter SettingsTo change the page of parameter:6 Repeat Steps 4 to 5 to change the desired parameter settings.7 To finish s

Seite 138

79OPERATING PROCEDURES: Parameter Settings22.8.1 Parameter tables[AR (AR measurement)]* Underlined options indicate factory settings.1 : STEP (AR meas

Seite 139 - GLOSSARY

80OPERATING PROCEDURES: Parameter Settings7 : AR THUMBNAILSelects whether or not to display the thumbnail screen of the measurement ring image during

Seite 140 - GLOSSARY:

81OPERATING PROCEDURES: Parameter Settings222 : LOW CONF LV (Display of confidence level)Setting of how to express the level of low confidence data wh

Seite 141

82OPERATING PROCEDURES: Parameter Settings[PRINT1 (Print setting 1)]* Underlined options indicate factory settings.31 : PRINT (printing)Selects the me

Seite 142

83OPERATING PROCEDURES: Parameter Settings238 : DATE FORMAT (date format)Selects the format of date.39 : PRINT COMMENTSelects whether or not to print

Seite 143

84OPERATING PROCEDURES: Parameter Settings44 : ERROR DATASelects whether or not to display and print erroneousdata obtained during AR measurement.When

Seite 144

85OPERATING PROCEDURES: Parameter Settings2[PRINT3 (Print setting 3)]* Underlined options indicate factory settings.51 : SE PRINT (printing of SE valu