Seitenansicht 117
104
MAINTENANCE: Error Messages and Countermeasures
ERR021
Error related to control signals for communication (input port)
Ensure that the interface cable is properly connected to the input port.
Also ensure that the communication parameters are properly set.
ERR022
Error related to control signals for communication (input port)
Ensure that the interface cable is properly connected to the input port.
Also ensure that the communication parameters are properly set.
ERR023
Error related to control signals for communication (input port)
Ensure that the interface cable is properly connected to the input port.
Also ensure that the communication parameters are properly set.
ERR024
Error related to control signals for communication (input port)
Ensure that the interface cable is properly connected to the input port.
Also ensure that the communication parameters are properly set.
ERR025
Error related to control signals for communication (input port)
Ensure that the interface cable is properly connected to the input port.
Also ensure that the communication parameters are properly set.
ERR026
Error related to communication data (output port)
Ensure that the interface cable is properly connected to the output port.
Also ensure that the communication parameters are properly set.
ERR027
Error related to communication data (output port)
Ensure that the interface cable is properly connected to the output port.
Also ensure that the communication parameters are properly set.
ERR028
Error related to communication data (output port)
Ensure that the interface cable is properly connected to the output port.
Also ensure that the communication parameters are properly set.
ERR031
Error related to the up-and-down tracking
Malfunction of the up-and-down motor, up-and-down sensor, electric circuit board, or break in
cables is probable.
Shut off the device and contact NIDEK or your authorized distributor.
ERR032
Error related to left-to-right tracking
Malfunction of the right-to-left motor, right-to-left sensor, electric circuit board, or break in cables
is probable.
Shut off the device and contact NIDEK or your authorized distributor.
ERR033
Error related to the back-and-forth tracking
Malfunction of the back-and-forth motor, back-and-forth sensor, electric circuit, or break in
cables is probable.
Shut off the device and contact NIDEK or your authorized distributor.
ERR034
Error related to the motorized chinrest
Malfunction for the chinrest motor, motorized chinrest sensor, or electric circuit board, or break
in cables is probable.
Shut off the device and contact NIDEK or your authorized distributor.
NO PAPER
Error related to the printer
If the printer is short of paper, refill paper.
If the printer cover is open, close it securely.
If the same error code is displayed even after the replacement of printer paper roll, malfunction of
the printer, electric circuit board, or break in cables is probable.
Shut off the device and contact NIDEK or your authorized distributor.
Error code Cause and countermeasure
Seitenansicht 117