OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements
1 Instruct the patient to look through the measuring
window and watch the center of a picture of a balloon
without straining as he or she sees it.
2 Manipulate the joystick to display the
patient’s eye on the screen.
By pulling the joystick backward, pushing it
forward, tilting it to the left and right, the main
body moves back, forth, to the right, and to
the left. By turning the upper part of the joy-
stick, the main body moves up and down.
Align the device to the patient’s eye with right,
left, up and down movements, and adjust the
focus with back and forth movements.
3 Perform alignment and focusing.
The methods of alignment and focusing vary
according to the 26. TRACKING SW parameter
setting.
*See “{ Auto-tracking mode and auto-shooting mode (
page 36) for details.
Perform alignment by placing the mire ring
reflected by the patient’s eye in the center of the
target.
Perform focusing according to the indication of
the on-screen focusing indicator.
• If the mire ring is not in the center of the pupil and the min. pupil mark is on the iris, turn off
the auto-tracking function and bring the min. pupil mark into the center of the pupil to
perform measurement, ignoring the mire ring.
• Auto-tracking or auto-shooting may not work on keratoconus or postoperative cornea.
In such a case, turn off the auto-tracking and auto-shooting functions.
Scenery picture with a balloon
ᵏᴷᴰǽǽǽǽǽᵈᴷᴰǽ
ᵐᴪǽᴮ
ᴫ
ᴯᴲǽᵏᴮǽǽᴴᴫᴶᴵǽᴮᴴᴰ
ᵀᴪǽᴭ
ᴫ
ᴴᴲǽᵏᴯǽǽᴴᴫᴲᴶǽǽᴵᴰ
ᴾǽǽǽǽǽᴯ
ᵏ
ᴥᴶᴦ
Mire ring
Target
Focusing indicator
KM measuring light